34 __________________________________________________________________ M210934EN-C

Set Analog Output Mode (mA/V)

The AMODE command can be used to change the analog output mode

of channels 1 and 2. When used without parameters it displays the

current settings and with parameters new settings can be set. The

available parameters are listed in Table 8 below.

Table 8

Analog Output Modes

Parameter

Output mode

1

0 … 20 mA

2

4 … 20 mA

4

0 … 5 V

5

0 … 10 V

6

LED alarm

7

Current alarm

8

Voltage alarm

Example:

amode 1 2

Ch1 output : 0 ... 20 mA

Ch2 output : 4 ... 20 mA

Calibrate Analog Output

Use the ACAL command to adjust analog output channels.
ACAL [channel]<cr>
where
channel = Analog output channel to be adjusted, 1 or 2.

After entering the ACAL command, use a calibrated current/voltage

meter to measure the output, and enter the values.

Example:

acal 1

Ch1 U1 ( V ) ? 1.001

Ch1 U2 ( V ) ? 9.011

acal 2

Ch2 U1 ( V ) ? 0.0988

Ch2 U2 ( V ) ? 0.8997