User's Guide ______________________________________________________________________

38 __________________________________________________________________ M210934EN-C

Set Output Interval

Use the INTV command to set or display the output interval.

INTV [n xxx]<cr>

where
n

= time interval = 0 ... 255

xxx

= time unit = "S", "MIN", or "H"

Example:

intv 1 min

Output interval: 1 MIN

The shortest output interval (with n = 0) is approximately one second due

to the internal measurement cycle.

Output Reading Once

Use the SEND command to output the reading once in STOP mode.

The output format depends on which parameters the transmitter can

output.

SEND [aaa]<cr>

where
aaa = Transmitter address (0 ... 255). Must be specified if the

transmitter is in POLL mode, and a line has not been opened

using the OPEN command.

Example:

send

T= 25.2 'C aw= 0.299 H2O= 19 ppm RS= 29.9 %