Technical data, Measured parameters, Chapter 7 – Vaisala MMT162 User Manual

Page 61

Advertising
background image

Chapter 7 ____________________________________________________________ Technical data

VAISALA _______________________________________________________________________ 59

CHAPTER 7

TECHNICAL DATA

This chapter provides the technical data of the product.

Measured Parameters

Water activity

Measurement range

0 ... 1 (@-40 ... +180 °C/ -40 ... +356 °F

)

Accuracy (including non-linearity, hysteresis and repeatability)

When calibrated against salt solutions (ASTM E104-85):

0 ... 0.9

± 0.02

0.9 ... 1.0

± 0.03

Response time in oil flow (typical)

with stainless steel filter

< 1 min. (dry-wet)

Sensor

HUMICAP®

Temperature

Measurement range

-40 ... +80 °C (-40 ... +176 °F)

Accuracy at +20 °C (+68 °F)

± 0.2 °C (± 32.36 °F)

Typical temperature dependence of

electronics

± 0.005 °C/°C (± 0.003 °F/°F)

Sensor

Pt100 RTD Class F0.1 IEC 60751

Advertising