One-point temperature adjustment, Using serial line – Vaisala MMT162 User Manual

Page 58

Advertising
background image

User's Guide ______________________________________________________________________

56 __________________________________________________________________ M210934EN-C

One-Point Temperature Adjustment

The MMT162 must be unpowered during stabilization which can take 30

minutes or more.

Using Serial Line

1.

Remove the probe filter and insert the probe into the reference

temperature.

2.

Enter the command CT and press ENTER:

CT<cr>

3.

Type C and press ENTER a few times to check if the reading is

stabilized. Let the reading stabilize, give the reference temperature

after the question mark and press ENTER three times.

Example:

ct

T : 22.03 Ref1 ?

T : 22.03 Ref1 ?

T : 22.03 Ref1 ? 25.0

Press any key when ready ...

T : 22.02 Ref2 ?

OK

4.

OK indicates that the calibration has succeeded.

5.

Take the probe out of the reference conditions and replace the

filter.