User's Guide ______________________________________________________________________

58 __________________________________________________________________ M210934EN-C

6.

Select 1-point adjustment and press Select.

7.

In adjustment mode, enter the correct reference value and then exit

by pressing Back.

8.

After you have exited the adjustment mode, MI70 confirms it by

displaying the message below.

Analog Output Adjustment

In the analog output calibration the analog output is forced to the

following values:
- current output: 2 mA and 18 mA

- voltage output: 10 % and 90 % of the range

Connect MMT162 to a calibrated current/voltage meter in order to

measure either current or voltage depending on the selected output type.

To calibrate analog output, see section Calibrate Analog Output on page

34.