Chapter 6 ___________________________________________________ Calibration and Adjustment

VAISALA _______________________________________________________________________ 53

6.

When the reading is stable, give the reference humidity after the

question mark and press ENTER.

crh

RH : 25.21 Ref1 ?

RH : 25.22 Ref1 ?

RH : 25.22 Ref1 ? 30.1

Press any key when ready ...

RH : 25.19 Ref2 ?

OK

7.

OK

indicates that the adjustment has succeeded and the new

calibration coefficients are calculated and stored.

8.

Take the probe out of the reference conditions and replace the

filter.

Using MI70

To adjust the relative humidity:

1.

Connect MI70 to MMT162.

2.

Turn MI70 on.

3.

On the MI70, go to MenuFunctionsAdjustments.