Chapter 5 ______________________________________________________________ Maintenance

VAISALA _______________________________________________________________________ 47

CHAPTER 5

MAINTENANCE

This chapter provides information that is needed in basic maintenance of

the product.

Cleaning

Clean the transmitter enclosure with a soft. lint-free cloth moistened with

mild detergent.

Clean the sensor before storing the MMT162 probe and before

calibration. For cleaning the the probe you need instrument air and

heptane (C

7

H

16

) liquid. If heptane is not available, you can also use diesel

oil or petrol. Dry with instrument air to prevent oxidation of the oil on the

sensor. The oxidation of the oil on the sensor can cause extended

response times or drifting.

1.

Blow the probe head (with filter) with instrument air to remove the

remains of oil.

2.

Immerse the probe head into heptane liquid and rinse out the oil

(for one minute maximum).

3.

Dry the probe head with instrument air. In case you are going to

calibrate the probe, remove the filter and dry the sensor with

instrument air. Check that the sensor looks clean.

NOTE

Do not immerse the sensor in any polar solvents, for example ketones

or alcohols.