Changing the filter, Changing the sensor, Error states – Vaisala MMT162 User Manual

Page 50

Advertising
background image

User's Guide ______________________________________________________________________

48 __________________________________________________________________ M210934EN-C

Changing the Filter

1.

Remove the filter from the probe.

2.

Install a new filter on the probe. When using the stainless steel

filter (for oil and fuel cell), take care to tighten the filter properly

(recommended force 130 Ncm).

New filters can be ordered from Vaisala, see section Spare Parts and

Accessories on page 61.

Changing the Sensor

If you need to change the sensor on your MMT162, send it to Vaisala

where the sensor will be changed for you.

Error States

MMT162 has the following states that indicate a problem with the probe:

- Analog 4 - 20 mA current output at 0 mA
- Analog voltage output at 0 V
- Serial line outputs asterisk characters (*** **)

The asterisk characters on the serial line may also appear during the start-

up procedure.

You can also check the error message via the serial interface by using the

ERRS command. In case of constant error, please contact Vaisala; see

section Technical Support on page 50.