IKA C 6000 isoperibol Package 2/12 User Manual
Page 13

13
Connecting the oxygen supply
Insert the O2 connection tube into the calorimeter
„IN“ port until it clicks home and connect the free
end to the pressure reducer, using the adapters
supplied if necessary. IKA
®
C 29. Removal entails
the same operations performed in the reverse
sequence.
Note: The O2 connection tube can be removed
only after it has been depressurised.
Screw the venting hose to the screw coupling
(8 AF) on the calorimeter and position the free
end under the extractor hood or connect it to a
gas washing device. Comply with the applicable
safety regulations in this respect.
O
2
connection tube IN
max. 4 MPa
Venting hose
IKA
®
C 29
Combustion gases are
hazardous to health,
therefore the venting hose must be
connected to a suitable gas cleaning system
or extraction system.
The venting hose discharges the combustion
gases from the decomposition vessel after every
combustion trial. When laying the venting hose
ensure that it is not crushed or kinked.
Connection to the mains power supply
Check that the available mains power supply
matches the mains power supply particulars listed
on the rating plate.
DANGER
Commissioning