IKA C 6000 isoperibol Package 2/12 User Manual

Page 52

Advertising
background image

52

Display

Description

What should be done?

What happens next?

No increase in temperature after
ignition.

This error message is generated,
if the increase in temperature
has not achieved the specifi ed
value within the fi rst minute af-
ter ignition.

Acknowledge the error message
to revert to the „Wait“ message.
The sample is combusted.
The weighted sample of the
sample may be insuffi cient.
The calorifi c value of the sample
is too small, use a combustion
aid.
The oxygen pressure is insuffi ci-
ent
The stirrer is not working.

The sample has not been com-
busted, despite the provision of
an ignition thread:
The ignition thread was blown
away from the sample
The oxygen pressure is insuffi ci-
ent

Sample and ignition thread were
not combusted:
Check the ignition system.

Measurement was cancelled by
the user.

This error message is triggered in
the Close, Run, Open conditions
if the user the presses the „Can-
cel“ button.

Acknowledge the error message
to revert to the „Wait“ message.

Cover not closed, or no decom-
position vessel is hung in the
calorimeter cover.

a. The cover did not reach the
lower cover contact.
b. The ignition contact did not
close.

Acknowledge the error message
to revert to the „Wait“ message.
a. Switch to the „Servicing“ tab.
Press the „Close“ button.
b. No decomposition vessel
hung in the cover
Ignition circuit open circuit.

Water circuit unstable

The stability criterion for the re-
gulation of the water circuits to
the working temperature was
not achieved within the specifi ed
period of time.

Acknowledge the error message
to revert to the „Wait“ message.
Check the water circuit
Check the heating
We recommend performing a
system test to localise more pre-
cisely the cause of the problem.

Troubleshooting