Display

Description

What should be done?

What happens next?

Controller alarm

The reliable regulation of the
outer vessel is a major precondi-
tion for good quality measured
values. Any error message from
the controller brings the current
measurement to an end and de-
mands a system test.
Further work using the calorime-
ter can be performed only once
the system test has been com-
pleted successfully.

Acknowledge the error message
in order to switch to the „System
Check“ status.
Press the „Restart“ button“.
The temperature converter and
the controller are reset and the
system test starts.
On successful completion the
status switches to „Wait“
If the system test was unsuccess-
ful, switch the C 6000 off and on
again.
If the system test is unsuccessful
again, please contact IKA

®

Cus-

tomer Service.

Troubleshooting