IKA C 6000 isoperibol Package 2/12 User Manual

Page 45

Advertising
background image

45

Empty:

The inner vessel can be emptied manually.

Stirrer:

Here the operation of the stirrer can be tested.

After the start the stirrer speed is displayed.

Cooling water:

After the start the current cooling water tempera-

ture at the entry into the calorimeter is displayed.

Cover up/down:

The cover can be opened and closed manually.

RFID:

After the start the decomposition vessel detec-

tion is activated and the RFID code of the decom-

position vessel can be read.

Click on „Stop“ to stop the function test.

Hardware details

In addition you can switch to the „Hardware de-

tails“ tab.

This tab shows you all the information about the

digital inputs and outputs, the temperature va-

lues on all the measurement channels, various

counters and statuses of accessories. You can-

not make any changes here. The information

displayed can be forwarded to Service Engineers

who have access to various settings.

Info

Here you can call up important information for

servicing purposes:

- Serial number of the calorimeter

- Software version

- PCB* number

- PCB* works number

- Number of ignitions

*PCB = printed circuit board

Help and servicing system