Display unit: this sets the display units for the

calorimetric results. Select them from the list.

The selected range of units applies to the display

of energy values [J], calorifi c values [J/g] and ther-

mal capacities [J/K].

Reference calorifi c value: Input here the ca-

lorifi c value of the substance used for calibration.

As a rule this substance is benzoic acid with a

calorifi c value certifi ed by the manufacturer.

„Measurement“ tab

Double determination: If this menu item is

activated, two measurements are basically com-

bined into a group.

The suffi x ЖRP_x is added onto the end of the

testing code. (X = 0, 1 or 2).

In the process, the fi rst two calorifi c values are

checked to ensure they fall within 120  J. If the

difference between them is greater than 120  J,

a warning appears that a third measurement is

required.

The third measurement must then fall within

120 J of one of the fi rst two measurements.

If the fi rst two measurements fall within 120 J of

each other, the next measurement creates a new

group.

Measurement

Other Settings

Select from the list:

The selected standard will be applied to all the

subsequent measurements.

Evaluated measurements will comply with this

standard: if a measurement has already been

evaluated, the evaluation procedure will be used

for any subsequent re-evaluation. Otherwise the

current setting for the evaluation procedure will

be used.

Measurements evaluated previously are not af-

fected by this change.