IKA C 6000 isoperibol Package 2/12 User Manual

Page 30

Advertising
background image

30

Securely screw the decomposition vessel to-

gether and hang it in the calorimeter cover, see

operating instruction decomposition vessel C

6010/6012.

The decomposition vessel must be positioned

centrally within the fi lling head of the inner cover.

When it is correctly positioned you can feel it click

home.

An instruction window now appears: „Check

that the decomposition vessel is securely closed“.

Click on the „Acknowledge“ button to acknow-

ledge the instruction.

When the decomposition vessel has been hung

in the calorimeter it is identifi ed by an RFID trans-

mitter (Rfi d: radio frequency identifi cation).

The decomposition vessel is fi tted with a trans-

ponder which is read as it passes over the sensor

area.

For this the decomposition vessel is passed over

the sensor area. A beep sounds when detection

has been successful.

You can then click on the „Start“ button.

Note: When working without automatic de-

composition vessel detection the „Start“ button

is active immediately the preparation has been

completed!

Starting the measurement

After the preparation for measurement, a mes-

sage appears requiring secure closure of the de-

composition vessel, despite there being no de-

composition vessel hung in the calorimeter cover.

You should acknowledge the message when you

have hung the decomposition vessel in the calori-

meter cover. If you accidentally press the „Start“

button before you have hung a decomposition

vessel in the calorimeter you will receive an error

message when you close the calorimeter cover.

Sensor area for decomposition

vessel detection

Working with the IKA

®

C 6000 global standards/isoperibol calorimeter