Complete combustion

To correctly determine the gross calorifi c value,
it is of fundamental importance that the sample
undergoes complete combustion. After each ex-
periment, the crucible and all solid residues must
be examined for any signs that combustion was
not complete.

In case of using substances tending to squirt it
is not possible to ensure complete combustion.
Substances with low fl ammability (substances

with a high content of mineral matter
and low-calorifi c substances) will often undergo
complete combustion only with the aid of com-
bustion capsules or combustion bags (C 10/ C 12
see „Accessories“). It is also possible to use liquid
combustion aids such as paraffi n oil,.
The combustion aids as well (for example the cot-
ton thread) must undergo complete combustion.
If unburned residues are left over, the experiment
must be repeated.

Adjustment (only IKA

®

C 6000 global standards)

If the device is to be operated in adiabatic func-
tion, prior adjustment in the respective tempera-
ture range (22°, 25° or 30°) is necessary.

Adjustment must be performed in the following cases:

When fi rst commissioning the calorimeter and

when its place of installation has been changed.

If the measuring times for adiabatic measure-

ments are regularly in excess of 15-20 minutes.

If adiabatic measurements are frequently abor-

ted because the time limits for the pre trial or
main trial have been exceeded.

Procedure for adjustment:

Set the function depending on the desired

temperature range (Adjustment_22, Adjust-
ment_25, Adjustment_30).

Prepare for a formal single measurement. An

empty crucible is used. The fi ctitious value „1“
is entered for the weighted sample.

Start the measurement.

Adjustment starts automatically and is comple-

ted within approx. 1 hour. After the adjustment
has been successfully completed the adjustment
value appears as an increase in temperature in
the measurement record for the measurement
that was performed.

After successful completion of adjustment the de-
vice automatically changes the mode of operation
to the corresponding adiabatic mode.

Adjustment_22

Adiabatic_22

Adjustment_25

Adiabatic_25

Adjustment_30

Adiabatic_30