Servicing the decomposition vessel: when you

wish to start using a decomposition vessel again

after servicing or repair, click on the „Servicing

the decomposition vessel“ button.

You will now see the name and RFID ID of the

selected decomposition vessel. Input the Service

Code for it and click on the „Renew“ button. You

will receive your service code once you have re-
gistered on the IKA

®

homepage.

Switch to the „Decomposition vessel“ tab and

click on the „Save“ button, then on the „Done“

button to revert to the main screen.

If you require settings different from the default

settings you can create those settings in a screen

with seven tabs.

„Evaluation“ tab

Default evaluation: NAfter a measurement has

been completed, as a rule the results are exten-

ded and corrected in accordance with the appli-

cable national or international standards.

Evaluation

Menu

Other settings

Servicing the decomposition vessel

New decomposition vessel: click on the „New

decomposition vessel“ button to assign a new

decomposition vessel to the calorimeter. Place

the new decomposition vessel over the sensor

face. The RFID tag is detected and displayed in

the „RFID fi eld (automatic decomposition vessel

detection“. Click on the „Serial number“ fi eld and

enter the engraved serial number for the new

decomposition vessel. Finally click on the „Add“

button.

Please note that the addition of a new de-

composition vessel cannot be undone.