IKA C 6000 isoperibol Package 2/12 User Manual

Page 27

Advertising
background image

27

A complete measurement can be performed in

accordance with the following steps without ma-

king further changes to the settings:

Click on the „Preparation“ button in order to

prepare for a new measurement. The following

screen appears:

With automatic decomposition vessel detection:

To make preparations for a valid measurement

you must input at least the weight of the sample.

Click on the „Weighted sample“ selection fi eld; a

virtual keyboard appears.

Preparing measurements

The term „measurements“ below refers to both
the measurements to calibrate the calorimeter
system (calibration measurements) and the actual
measurements for determining the calorifi c value.
The difference lies mainly in the evaluation (see

the section „Working with the device“ and „Set-
tings“), whilst preparation and performance make
virtually no difference.

The following preparations must be performed

in order to prepare the system to take a

measurement:

1.With the decomposition vessel open, attach a

cotton thread to the centre of the ignition wire

using a loop.

2. Weigh out the substance directly into the

crucible with an accuracy of 0.1 mg. If necessary

put some distilled water or a solution into the

decomposition vessel.

The maximum acceptable weight of the sample

that is added is restricted to a value of 5 g.

To prolong the life of wearing parts (o-rings,

seals, etc.) we recommend that you always

work with a water trap.

As a rule the weighted sample must be

selected in such a way that the temperature

increase during the measurement is below 5 K

and comes close to the temperature increase

of the calibration (max. extra energy: 40,000

J). Failure to observe these instructions could

result in damage to the decomposition vessel.

If the maximum energy input is exceeded, we

recommend that the calorimeter is sent back

for repair.

When working with unknown substances, at

fi rst select only very small weighted samples

(approx. 0.25 g) to determine the energy

potential. If you are burning unknown samples,

leave the room or keep well away from the

calorimeter.

If substances such as distilled water or

solutions are added to the decomposition

vessel for combustion tests, you must use

exactly the same amount of those substances

during calibration.

Refer also to the operating instructions for the

C 6010/C 6012 decomposition vessel.

Working with the IKA

®

C 6000 global standards/isoperibol calorimeter

Advertising