29

You can input values between 0 and 20,000

Joules for „External energy 2“. You can also select

specifi c types of combustion aids from the list.

In such cases you must enter the weight of this

combustion aid so that the external energy can

be calculated.

You can also load these weights directly from a

balance: The most fl exible way is to click on the

relevant „Balance“ button. The value is then loa-

ded to the right place.

If the balance cannot receive commands, still

click on the „Balance“ button and then press the

„Transfer“ button for the balance.

Simulated measurement:

If the measurement is marked as simulation, the

increase in temperature can be set by the user.

This means that the results can be verifi ed later.

You can also load weights without clicking on the

„Balance“ button:

If the „Preparation for measurement“ screen has

not yet been opened, open it. The weights are in-

put in the following sequence:

Weight of combustion aid 2 (if defi ned)

Weight of the sample

Weight of combustion aid 3 (if defi ned).

After all inputs have been performed in the „Pre-

paration“ screen, click on OK; the display along-

side will appear.

Note: When working without automatic de-

composition vessel detection the „Start“ button

is active immediately the preparation has been

completed!

Calorimetric procedure without automatic decomposition vessel detection

Firstly the automatic decomposition vessel detec-

tion must be deactivated (see „New decompositi-

on vessel“ tab, page 37).

The procedure is the same as the standard proce-

dure with the following exceptions:

Preparation for measurement

For preparation for measurement you must select

a decomposition vessel from the list.

Charging the decomposition vessel

Please refer to the operating instructions for the

C 6010/ C 6012 decomposition vessels.

Working with the IKA

®

C 6000 global standards/isoperibol calorimeter

Advertising