IKA C 6000 isoperibol Package 2/12 User Manual

Page 36

Advertising
background image

36

„Calibration“ tab

You can select each calibration individually and

click on the „Select“ button, or double click on

the calibration to reference this calibration in the

calculation of the C mean value (or to exclude it

if it had previously been selected). Then switch to

the „Statistics“ tab.

„Statistics“ tab

Average value, standard deviation and maximum

deviation relate to the calibration selected. Click

on the button „Load“ to load the average value

of the calibrations as the new C value.

You can also press the „Set“ button to input the

new C value manually. Press the „Confi rm“ but-

ton to load the value.

Press the „Print statistics“ button to print the

complete statistics on a connected printer. Press

the „Delete statistics“ button to delete the selec-

tion from the statistics that are referred to during

calibration.

Click on the „Save“ button, to use all the chan-

ges for the selected decomposition vessel and

the selected function. If you click on the „Undo“

button, you can completely undo these changes.

If you press „Done“ you save the inputs, exit this

menu, and revert to the overview of the measu-

rements.

If you press „Save“ you remain in this menu and

can select other actions (such as „Check chart“).

Menu

If you wish to change the function for subsequent

measurements, click on the „Change“ button to

start the system check for the new function.

Alternatively you can select a decomposition ves-

sel from the updated list of decomposition ves-

sels.

Advertising