Chapter 3 SQS-142 Software

3-18

Sources Tab

Output Name: A meaningful name assigned to SQM-142 outputs 1 and 2 in single card
systems; 1 to 4 in two card systems. For easy display, keep the name to less than 8
characters.

Full Scale: The input voltage required by the deposition source power supply to
produce 100% output power. Positive or negative full scale values are possible.

Completion Timer: If Complete Signal is checked, the process will halt if a move
complete signal is not received within this timeout period. If Complete Signal is not
checked, the process waits for this fixed time period before starting a layer.

Complete Signal: Check this box if your source index sends a signal designating that
the move is complete and preconditioning can begin.

Test Output: Select an output, then click Send to set the SQM-142 card output to its
Full Scale voltage. Click Clear to return the selected output to 0 volts.