TM-350/400 Operating Manual

4.3.2 Computer Interface Address

The TM-350/400 allows for computer interfaces between 1 and 32 for RS-485 and
IEEE-488 communications. As shipped the Monitor is set up for an interface
address of 1. To reconfigure the monitor for any other interface address, switches
4, 5, 6, 7, and 8, respectively, of S1 need to be set as shown in

Table 4-2

.

Table 4-2 Reconfiguring Monitor for Other Interface Address

Interface Address

S/SEL 4

S/SEL 5

S/SEL 6

S/SEL 7

S/SEL 8

1

OFF OFF OFF OFF OFF

2 ON

OFF

OFF

OFF

OFF

3

OFF ON OFF OFF OFF

4 ON

ON

OFF

OFF

OFF

5 OFF

OFF

ON

OFF

OFF

6

ON OFF ON OFF OFF

7 OFF

ON

ON

OFF

OFF

8

ON ON ON OFF OFF

9

OFF OFF OFF ON OFF

10 ON

OFF

OFF

ON

OFF

11

OFF ON OFF ON OFF

12 ON

ON

OFF

ON

OFF

13 OFF

OFF

ON

ON

OFF

14

ON OFF ON ON OFF

15 OFF

ON

ON

ON

OFF

16

ON ON ON ON OFF

17

OFF OFF OFF OFF ON

18 ON

OFF

OFF

OFF

ON

19

OFF ON OFF OFF ON

20 ON

ON

OFF

OFF

ON

21 OFF

OFF

ON

OFF

ON

22

ON OFF ON OFF ON

23 OFF

ON

ON

OFF

ON

24

ON ON ON OFF ON

25

OFF OFF OFF ON ON