INFICON PSG55x ATM to Medium Vacuum Gauge User Manual

Page 11

Advertising
background image

tina60e1-a (2011-04)

11

Output impedance

2 × 4.7

Ω, short circuit-proof

Load impedance

≥10 k

Ω

Response time

<10 ms

Gauge identification

FCC 68 (+0.61 … +10 V)


27 k

Ω

HV adjustment

at <10

-5

mbar

Solid state relays

switching functions SP1,
SP2, ATM

Setting range (N

2

) 5.0×10

-5

… 1000 mbar

Hysteresis

1)

10%

of

threshold

Switching characteristics

1)

Low Trip Point

Contact rating

<30 VAC/DC, ≤0.3 A
resistive

closed LED

lit

solid

open LED

off

Switching time

<30 ms


Mechanical relays

switching functions SP1,
SP2, ATM

Setting range (N

2

) 5.0×10

-5

… 1000 mbar

Hysteresis

1)

10%

of

threshold

Switching characteristics

1)

Low Trip Point

Type

1 floating contact (n.o.) per
switching function

Contact rating

<30 VAC/DC, ≤1 A
resistive

closed LED

lit

solid

open LED

off

Switching time

<30 ms

Diagnostic port

Jack connector 2.5 mm,
3-pin

1

)

The hysteresis and the switching characteristics can be

programmed via the serial interface or the diagnostic port.

Advertising