4.2

Gas Type Dependence

The measurement value is gas dependent. The pressure reading
applies to dry air, O

2

, CO and N

2

. For other gases, it has to be

corrected (→ "Technical Data").
If the gauge is operated with an INFICON controller, a calibration
factor for correction of the actual reading can be applied (→

of

the corresponding controller).

4.3

Switching Functions SP1, SP2

The two switching functions can be set to any pressure within
the measurement range of the gauge. A solid state relay is
provided for each switching function.

The current threshold setting
• can be read / written via the diagnostic port
• is output at the measurement signal output instead of the

pressure signal, can be measured with a voltmeter, and is
displayed on the LCD display after the <SP1> or <SP2>
button is pressed

• can be read / written via the DeviceNet, Profibus and RS485

interface.

Switching characteristics and hysteresis

The switching characteristics and the hysteresis of each set
point can be programmed (

→  41).

Low Trip Point

(default)

If the pressure in the vacuum system is lower than the setpoint,
the corresponding LED (<SP1> or <SP2>) is lit solid and the
corresponding relay is closed.