2 gas type dependence, C × indicated pressure, Calibration factors – INFICON PSG55x ATM to Medium Vacuum Gauge User Manual

Page 23: Indicated pressure (gauge calibrated for air), Valid for pirani pressure range below 1 mbar, P (mbar) p, Mbar), Air, o, Co, n

Advertising
background image

tina60e1-a (2011-04)

23

2.2

Gas Type Dependence

Indicated pressure (gauge calibrated for air)

10

2

8
6
4

2

10

1

8

6
4

2

10

0

8

6
4

2

10

–1

8

6
4

2

10

–2

8

6
4

2

10

–3

10

–3

2

4

6

10

–2

2

4

6

10

–1

2

4

6

10

0

2

4

6

10

1

2

4

6

10

2

H

2

He

Ne

Xe

Kr

Ar

CO

2

Freon 12

Air

O

2

CO

N

2

p (mbar)

p

eff

(mbar)

Water vapor

Calibration factors

valid for Pirani pressure range below 1 mbar

p

eff

= C × indicated pressure

Gas type

Calibration

factor C

Gas

type Calibration

factor C

He
Ne

Ar
Kr

Xe

0.8
1.4
1.7
2.4
3.0

H

2

air, O

2

, CO, N

2

CO

2

water vapor

Freon 12

0.5
1.0
0.9
0.5
0.7

Advertising