4.6 Profibus

Operation

Caution

Caution: data transmission errors
The attempt to operate the gauge with the RS232C
interface causes data transmission errors.
This gauge must not be operated with the RS232C
interface.

For operating the gauge via Profibus, prior installation of the
device specific GSD file is required on the bus master side. This
file can be downloaded from our website.

Node Address Setting

For unambiguous identification of the gauge in a Profibus en-
vironment, a node address is required.

4

2

6

0

8

E

A

C

LSD

MSD

ADDRESS

4

2

6

0

8

E

A

C

Node address 0 … 125

dec

The node address is set in hexadecimal form
(00 … 7D

hex

) via the <MSD> and <LSD>

switches. It can not be defined via Profibus.

Node address >7D

hex

(>125

dec

)

The gauge starts with the node address 126

dec

. The address can

now be set via Profibus ("Set slave address",

→ [5]). Addi-

tionally, via the attribute "NO_ADD_CHG" can be defined, if fur-
ther changes of the node address are permissible.
The values of the nude address and the attribute are stored non-
volatile. To change these stored values, start the gauge with a
node address <126

dec

. The stored values of the nude address

and the attribute are deleted.