INFICON PSG55x ATM to Medium Vacuum Gauge User Manual
Page 13
tina60e1-a (2011-04)
13
RS232C / RS485 interface
Transmission rate
Data format
57600 baud (default)
binary
8 data bits
one stop bit
no parity bit
no handshake
→ "Power Connection"
For further information on the RS232C / RS485C interface
→ [3].
DeviceNet interface
Specification, data format,
communication protocol
→ [6]
Interface, physical
CAN bus
Data rate (adjustable via <RATE>
switch)
125 kBaud
250 kBaud
500 kBaud (default)
<P> (125 kBaud,
250 kBaud, 500 kBaud
programmable via
DeviceNet,
→ [4])
Node address (MAC ID)
(Adjustable via <ADDRESS>,
<MSD>, <LSD> switches)
0 … 63dec (63dec default)
<P> (0 … 63 programmable
via DeviceNet,
→ [4])
DeviceNet connector
Micro-Style, 5-pin, male
Cable
shielded, special DeviceNet
cable, 5 conductors
→ 32, → [7]
Cable length, system wiring
according to DeviceNet
specifications,
→ [6], [7]
For further information on the DeviceNet interface
→ [4]