INFICON PSG55x ATM to Medium Vacuum Gauge User Manual
Page 41
tina60e1-a (2011-04)
41
4.3.1
Adjusting the Setpoints SP1, SP2
The switching characteristics and the hysteresis can
only be programmed via
• the diagnostic port (→ [3])
• the DeviceNet, Profibus and RS485 interface
(
→ [3], [4], [5]).
The thresholds of the setpoints can be adjusted via
• the buttons on the gauge
• the diagnostic port (→ [3])
• the DeviceNet, Profibus and RS485 interface
(
→ [3], [4], [5]).
If both a High Trip Point and a Low Trip Point are
assigned to a setpoint, Low Trip Point only can be
adjusted via the corresponding button on the gauge.
DANGER
DANGER: malfunction
If processes are controlled via the signal output,
keep in mind that by pushing an <SP> button the
measurement signal is suppressed and the cor-
responding threshold value is output instead. This
can cause malfunctions.
Push the <SP> button only if you are sure that no
damages can arise from a malfunction.