P = 10, U = c + log, Measurement range 1.2 … 8.5 v – INFICON PSG55x ATM to Medium Vacuum Gauge User Manual

Page 19: Valid in the range 5×10, Mbar <p< 1000 mbar, Output signal u [v] pressure p

Advertising
background image

tina60e1-a (2011-04)

19

Measurement range 1.2 … 8.5 V

Output signal U [V]

Pressure p

1E+05

1E+04

1E–03

1E–04

1E–05

0.0 0.5 1.0 1.5

2.5

3.5

4.5

5.5

6.5

7.5

8.5

2.0

3.0

4.0

5.0

6.0

7.0

8.0

9.0

1.2

mbar

Pa

Torr

1E+03

1E+02

1E+01

1E+00

1E–01

1E–02

Er

ro

r

p = 10

(U-c)

U = c + log

10

p

valid in the range

5×10

-5

mbar <p< 1000 mbar

U p

c

U p

c

[V] [mbar]

5.5

[V] [micron]

2.625

[V] [µbar]

2.5

[V] [Pa]

3.5

[V] [Torr]

5.625 [V] [kPa]

6.5

[V] [mTorr]

2.625

where p pressure

U output signal
c constant (pressure unit dependent)

Advertising