2 power connection – INFICON PSG55x ATM to Medium Vacuum Gauge User Manual

Page 27

Advertising
background image

tina60e1-a (2011-04)

27

3.2 Power

Connection

Make sure the vacuum connection is properly made

(

→  24).

DANGER

The gauge may only be connected to power sup-
plies, instruments or control devices that conform
to the requirements of a grounded protective extra-
low voltage (SELV) and limited power source
(LPS), Class 2. The connection to the gauge has to
be fused.

5)

Ground loops, differences of potential, or EMC problems

may affect the measurement signal. For optimum signal
quality, please do observe the following notes:

• Connect the cable shield to ground on one side via

the connector housing. Do not connect the other side
of the shield.

• Connect the supply common with protective ground

directly at the power supply.

• Use differential measurement input (signal common

and supply common conducted separately).

• Potential difference between supply common and

housing ≤18 V (overvoltage protection).

5)

INFICON controllers fulfill these requirements.