3 troubleshooting – INFICON PSG55x ATM to Medium Vacuum Gauge User Manual

Page 54

Advertising
background image

54

tina60e1-a (2011-04)

6.3 Troubleshooting

In case of an error, it may be helpful to just turn off the

mains supply and turn it on again after 5 s.

Problem Possible

cause Correction

Output signal per-
manently

≈0V

Sensor cable defective or not
correctly connected

Check the sensor cable

No supply voltage

Turn on the power supply

Error

Remedy the error

Gauge in an undefined status

Turn the gauge off and on
again after 5 s (reset)

FAIL PIR1

Pirani sensor defective Replace

the

sensor

(

→  53)

Electronics unit not correctly
mounted on sensor

Check the connections
(electronics – sensor)

FAIL EEPROM

EEPROM error

Turn the gauge off and on
again after 5 s (reset)

Replace

the

gauge

FAIL SENSOR

Electronics unit not compatible
with the sensor

Replace the sensor
(

→  53)

Replace

the

gauge