Press the <ADJ> button with a pin (max. ø1.1 mm) and the

ATM adjustment is carried out: The Pirani sensor is ad-
justed to 1000 mbar by default.

ATM

HV

q

Evacuate the vacuum system to p << 10

-5

mbar and wait at

least 2 minutes.

r

Press the <ADJ> button with a pin and the HV adjustment

is carried out: The gauge is adjusted to 5×10

-5

mbar

(default).

HV adjustment to another pressure

→ [3].

If the pressure value 4.99×10

-5

mbar is output at the

measurement value output or on the LCD display, the
adjustment has been successful. Otherwise, repeat the
adjustment procedure.