tina31e1 (2004-05) HPG400 v1.om

11

DeviceNet connector

Micro-Style, 5 pins, male

Cable

Shielded, special DeviceNet cable,
5 conductors (

→ 19 and [4])

Cable length, system wiring

According to DeviceNet specifications
(

→ [4], [6])

For HPG400-SP gauges with part numbers:

353-525
353-526

Fieldbus name

Profibus

Standard applied

→ [7]

Communication protocol, data format

→ [2], [7]

Interface, physical

RS485

Data rate

≤12 MBaud (→ [2])

Node address

Local:
(Adjustable via hexadecimal
"ADDRESS", "MSD", "LSD"
switches)

Via Profibus:
(hexadecimal "ADDRESS" switches
set to >7d

hex

(>125

dec

))

00 … 7D

hex

(0 … 125

dec

)

00 … 7D

hex

(0 … 125

dec

)

Profibus connection

D-Sub, 9 pins, female

Cable

Shielded, special Profibus cable
(

→ 20 and [5])

Cable length, system wiring

According to Profibus specifications
(

→ [5], [7])

Materials exposed to vacuum

Housing, supports, screens
Feedthroughs
Insulator
Cathode
Cathode holder
Pirani element

stainless steel
NiFe, nickel plated
glass
iridium, yttrium oxide (Y

2

O

3

)

molybdenum
tungsten, copper

Internal volume

353-520, 353-521
353-525, 353-527

353-522, 353-523
353-526, 353-528

≤20 cm

3

≤20 cm

3

≤30 cm

3

≤30 cm

3

Pressure max.

5 bar (absolut)
(only for Inert gases and
temperatures <100 °C).

Admissible temperatures

Storage

Operation

Bakeout

-20 … 70 °C

0 … 50 °C

+150 °C (without electronics unit)

Relative humidity
(year's mean / during 60 days)

≤65 / 85% (no condensation)

Use

indoors only
altitude up to 2000 m NN

Type of protection

IP 30

Profibus interface
(HPG400-SP)

Vacuum

Environment