4 Operation

The HPG400 consists of two separate measuring systems (high pressure (HP) hot
cathode and Pirani).

The HP hot cathode measuring system is based on the electrode arrangement
shown in the figure below, which grants sensitivity, linearity, and stability even at
high pressures.

The measuring principle of this system is based on gas ionization. Electrons emit-
ted by the hot cathode (F) ionize a number of molecules proportional to the pres-
sure in the measuring chamber. The ion collector (IC) collects the thus generated
ion current I

+

and feeds it to the electrometer amplifier of the measuring instrument.

The ion current is dependent on the emission current I

e

, the gas type, and the gas

pressure p according to the following relationship:

I

+

= I

e

× p × C

Factor C represents the sensitivity of the gauge. It is generally specified for N

2

.

The lower measurement limit is 2×10

-6

mbar.

In order for the entire range of 2×10

-6

mbar … 1 mbar to be covered, the emission

current is continually increased from 4 µA at 1 mbar to 130 µA at 2×10

-6

mbar (no

transients due to switching of the emission current).