B: gas type dependence – INFICON HPG400 ATM to High-Vacuum Gauge User Manual

Page 46

Advertising
background image

46

tina31e1 (2004-05) HPG400 v1.om

Output signal U

[V]

[mbar]

Pressure p

[Torr]

[Pa]

0 … 0.5

Sensor error

0.5 … 1.5

Underrange

1.5

2.5

3.5

4.5

5.5

6.5

7.5

1.0Ч10

-6

1.0Ч10

-5

1.0Ч10

-4

1.0Ч10

-3

1.0Ч10

-2

1.0Ч10

-1

1

7.5Ч10

-7

7.5Ч10

-6

7.5Ч10

-5

7.5Ч10

-4

7.5Ч10

-3

7.5Ч10

-2

7.5Ч10

-1

1.0Ч10

-4

1.0Ч10

-3

1.0Ч10

-2

1.0Ч10

-1

1

10

100

7.5 … 8.0

Overrange

8.0 … 8.5

Underrange

8.5

8.75

9.0

9.25

9.5

9.75

1.0Ч10

-2

1.0Ч10

-1

1

10

100

1000

7.5Ч10

-3

7.5Ч10

-2

7.5Ч10

-1

7.5

75

750

1

10

100

1000

10000

100000

9.75 … 10.2

Overrange

For gases other than air, the pressure can be determined by a simple conversion:

p

eff

= K × pressure indicated

Where

Gas type

K (mean values)

Air (N

2

, O

2

)

1.0

Xe

0.4

Kr

0.5

Ar

0.8

H

2

2.4

Ne

4.1

He

5.9

Conversion table
output signal - pressure

B:

Gas Type Dependence

Hot cathode range