Appendix, A: relationship output signal – pressure – INFICON HPG400 ATM to High-Vacuum Gauge User Manual

Page 45

Advertising
background image

tina31e1 (2004-05) HPG400 v1.om

45

Appendix

over

range

1E+03

1E+02

1E+01

1E+00

0.0 0.5 1.0 1.5

2.5

3.5

4.5

5.5

6.5

7.5

8.5

9.5

2.0

3.0

4.0

5.0

6.0

7.0

9.0

1E–01

1E–02

1E–03

1E–04

1E–05

1E–06

8.0

10.2

10.0

Measuring signal U[V]

Pressure p [mbar]

Hot cathode range and Pirani range require different conversion formulae:

p = 10

U - c1

U = c1 + log p

p = 10

(4 × (U - c2))

U = c2 + 0.25 log p

Where

U

p

c1

c2

[V] [mbar]

7.5

9

[V] [Torr]

7.625

9.031

[V] [Micron]

4.625

8.281

[V] [Pa]

5.5

8.5

Switch position

<P

↔ HP>

Threshold

0 or 1
2 or 3
4 or 5
6 or 7
8 or 9

1 mbar (factory setting)

5×10

–1

mbar

2×10

–1

mbar

1×10

–1

mbar

5×10

–2

mbar

A:

Relationship Output
Signal – Pressure

Conversion curve

Conversion formula
hot cathode range

Conversion formula
Pirani range

Changeover threshold
hot cathode

↔ Pirani