INFICON HPG400 ATM to High-Vacuum Gauge User Manual

Page 9

Advertising
background image

tina31e1 (2004-05) HPG400 v1.om

9

For gauges with part numbers

353-521
353-523

Display panel

Dimensions

Parameters

Pressure units

LCD matrix, 32×16 pixels,
with background illumination
green = normal operation
red = error/warning

→ 27)

16.0 mm × 11.2 mm

Pressure, pressure unit, status,
error messages

mbar (factory setting), Torr, Pa
(selecting the pressure unit

→ 28)

DANGER

The gauge must only be connected to power supplies, instruments or
control devices that conform to the requirements of a grounded extra-
low voltage (SELV-E according to EN 61010). The connection to the
gauge has to be fused (INFICON controllers fulfill these requirements).

Operating voltage at the gauge

+24 VDC (20 … 28 VDC)

1)

ripple max. 2 V

pp

Power consumption

Standard
Emission start (<200 ms)

≤0.5 A
≤1.4 A

Power consumption

HPG400, HPG400-SD
HPG400-SP

≤16 W
≤18 W

Fuse necessary

1.25 AT

HPG400-SD requires an additional, separate power supply for the
DeviceNet interface (

→ 19).

Supply voltage at the DeviceNet
connector (Pin 2 and Pin 3)

+24 VDC (+11 … 25 VDC)

Power consumption

<2 W

The gauge is protected against reversed polarity of the supply voltage.

1)

Measured at sensor cable connector (consider the voltage drop as function of
the sensor cable length).

Display (

HPG400)

Power supply