INFICON HPG400 ATM to High-Vacuum Gauge User Manual

Page 32

Advertising
background image

32

tina31e1 (2004-05) HPG400 v1.om

The admissible data rate depends on a number of factors such as system pa-
rameters and cable length (

→ [4], [6]) ]). It can be set on the gauge or pro-

grammed via DeviceNet.

By means of the "RATE" switch, the data rate can be set to 125
("1"), 250 ("2") or 500 kBaud ("5").

If the switch is in any of the "P" positions, the data rate is pro-
grammable via DeviceNet (

→ [1]).

Two lights (LEDs) on the gauge inform on the gauge status and the current
DeviceNet status.

Light status

Description

Dark

No supply

Flashing

red/green

Selftest

Green

Normal operation

Red

Non recoverable error

Light status

Description

Dark

Gauge not online:

− Selftest not yet concluded
− No supply, → "STATUS MOD" light

Flashing

green

Gauge online but no communication:

− Selftest concluded but no communication to other nodes

established

− Gauge not assigned to any master

Green

Gauge online; necessary connections established

Flashing red

One or several input/output connections in "timed out" status

Red

Communication error. The gauge has detected an error that im-
pedes communication via the network (e.g. two identical node ad-
dresses (MAC IC) or "Bus-off")

The gauge is connected to the DeviceNet system via the 5-pin DeviceNet con-
nector(

→ 19).

4.6.2.3 Data Rate Setting

4.6.3 Status Lights

"STATUS MOD"
(gauge status)

"STATUS NET"
(network status)

Electrical connections