2 electrical connection – INFICON HPG400 ATM to High-Vacuum Gauge User Manual
Page 16
16
tina31e1 (2004-05) HPG400 v1.om
Slide the electronics unit in to the mechanical stop and lock it with the hexa-
gon socket set screw (1).
The electronics unit is now installed.
If the gauge is used with an INFICON VGC4XX controller, a corresponding sensor
cable is required (
→ 43). The sensor cable permits supplying the gauge with
power, transmitting measurement values and gauge statuses, and making pa-
rameter settings.
Caution
Caution: data transmission errors
If the gauge is operated with the INFICON VGC4XX Vacuum Gauge
Controller (RS232C) and a fieldbus interface at the same time, data
transmission errors may occur.
The gauge must not be operated with an INFICON VGC4XX controller
and DeviceNet or Profibus at the same time.
• Sensor cable (→ [10] INFICON sales literature)
Plug the sensor connector into the gauge and secure it with the locking
screws.
Connect the other end of the sensor cable to the INFICON controller and
secure it.
The gauge can now be operated with the VGC4XX controller.
3.2 Electrical Connection
3.2.1 Use With INFICON
VGC4XX Vacuum Gauge
Controller
Required material
Procedure