tina31e1 (2004-05) HPG400 v1.om

Electrical connection

Pin 1 Relay Switching function A, COM contact
Pin 2 Signal output (measuring signal)

0 … +10 V

Pin 3 Threshold value (Setpoint) A

0 … +10 V

Pin 4 Relay Switching function A, N.O. contact
Pin 5 Supply common, GND
Pin 6 Threshold value (Setpoint) B

0 … +10 V

Pin 7 not connected

Pin 8 Supply voltage

+24 V

Pin 9 Relay Switching function B, COM contact
Pin 10 Gauge identification
Pin 11 Relay Switching function B, N.O. contact
Pin 12 Signal common, GND
Pin 13 RS232, TxD
Pin 14 RS232, RxD
Pin 15 Shielding, housing, GND

HPG400-SD, HPG400-SP

8

9

1

15

D-Sub, 15 pins

female

soldering side

SP A

SP B

TxD

RxD

+U

b

56 k

Measuring

signal

Threshold value, SP A

Threshold value, SP B

3

6

1

4

9

11

13

14

7

8

2

12

5

15

1.25 AT

24V

Identification

RS232

10

WARNING

The supply common (Pin 5) and the shielding (Pin 15) must be
connected at the supply unit with protective ground.

Incorrect connection, incorrect polarity or inadmissible supply
voltages can damage the gauge.

For cable lengths up to 5 m (0.34 mm

2

conductor cross-section) the out-

put signal can be measured directly between the positive signal output
(Pin 2) and supply common GND (Pin 5) without loss of accuracy. At
greater cable lengths, differential measurement between signal output
(Pin 2) and signal common (Pin 12) is recommended.

Reassemble the cable connector.

On the other cable end, terminate the cable according to the requirements

of the gauge controller you are using.

Sensor cable connection
HPG400-SD, HPG400-SP