2 use with other controllers – INFICON HPG400 ATM to High-Vacuum Gauge User Manual

Page 17

Advertising
background image

tina31e1 (2004-05) HPG400 v1.om

17

The gauge can also be operated with other controllers.

Especially the fieldbus versions HPG400-SD (DeviceNet) and HPG400-SP
(Profibus) are usually operated as part of a network, controlled by a master or bus
controller. In such cases, the control system has to be operated with the appropri-
ate software and communication protocol (

→ [1] and [2]).

For reasons of compatibility, the expression "sensor cable" is used for all
HPG400 versions in this document, although the pressure reading of the
gauges with fieldbus interface (HPG400-SD or HPG400-SP) is normally
transmitted via DeviceNet or Profibus.

The sensor cable is required for supplying all HPG400 types with power.
In connection with the gauges with fieldbus interface (HPG400-SD and
HPG400-SP), it also permits access to the relay contacts of the switch-
ing functions (

→ 18, 34).

The application and length of the sensor cable have to be considered when deter-
mining the number and cross sections of the conductors (

→ 10).

Open the cable connector (D-Sub, 15 pins, female).

Prepare the cable and solder/crimp it to the connector as indicated in the

diagram of the gauge used:

Electrical connection

Pin 2 Signal output (measuring signal) 0 … +10 V
Pin 5 Supply voltage common, GND
Pin 8 Supply voltage +24 VDC
Pin 10 Gauge identifcation
Pin 12 Signal common, GND
Pin 13 RS232C, TxD
Pin 14 RS232C, RxD
Pin 15 Shielding, housing, GND

Pins 1, 3, 4, 6, 7, 9 and 11
are not connected internally.

8

9

1

15

TxD

HPG400

56k

RxD

Measuring signal

+Ub

13

14

2

12

8

5

15

+

1.25AT

Identification

10

RS232C

+

24 V

D-Sub, 15 pins

female

soldering side

3.2.2 Use With Other

Controllers

3.2.2.1 Making a Sensor Cable

Cable type

Procedure

Sensor cable connection
HPG400

Advertising