4 display (hpg400) – INFICON HPG400 ATM to High-Vacuum Gauge User Manual

Page 27

Advertising
background image

tina31e1 (2004-05) HPG400 v1.om

27

The gauges with part numbers

353-501 and
353-503

have a built-in two-line display with an LCD matrix of 32×16 pixels. The first line
shows the pressure in normal operation, the second line the pressure unit, the
function and operational status and possible error messages.

The background illumination is usually green, in the event of an error, it changes to
red.

Pressure reading

Pressure unit

Status

Pressure reading in mbar (factory setting), Torr or Pa.

The pressure unit can be changed via RS232C interface
(

→ 28).

E

A

=

=

Emission of HP hot cathode on

Adjustment at atmospheric
pressure in progress

ok

=

Normal operation, no error
(green background illumination)

FAIL lonG 5
FAIL lonG 8
FAIL lonG 9

=

=

=

Pirani sensor warning

Hot cathode sensor warning

Pirani sensor error

(red background illumination)

Internal data connection failure
(red background illumination)

4.4 Display (HPG400)

Pressure display

Function display (status)

Error display