11



LH 46

LH 85

Mains power connection 230/60 V/Hz

207-253V/60Hz

207-253V/60Hz

Current draw

(at 230 V) A

15

16

Working temperature range

°C

-45 ... 250

-80 ... 250

Temperature stability

K

±0,01 ... ±0,05

±0,01 ... ±0,05

Cooling capacity

pump pressure stage 1

pump pressure stage 5

(bath liquid: Thermal / Ethanol)

°C
kW
kW

200 / 20 -20 -40
2.5 / 2.35 0.72 0.13
2.3 / 2.28 0.67 0.03

100 / 20 -20 -40 -60 -80
1.35 / 1.02 0.85 0.76 0.52 0.01
1.2 / 0.9 0.71 0.62 0.49 0.01

Cooling compressor

cooling with

1-stage
air / water

2-stage
air / water

Refrigerant

R404A

R23 / R404A

Cooling water : Flow rate

at 15 °C inlet temperature

l/h

360

198

Heater wattage

kW

1,8

1,8

Pressure pump, adjustable
pressure min ... max. at 0 liter

discharge, max. at 0 bar

grade

bar

l/min

1

0.5

16

2

0.7

19

3

1.1

23

4

1.3

26

5

1,7

30

maximum permissible viscosity is 50 mm² / s.

Noise level, 1 m distance

dBA

-----

-----

Ambient temperature

°C

5 ... 40

5 ... 40

Filling volume

Heat exchanger

Reservoir

liters

liters

2.5

1.2 … 3.4

2.5

1.2 … 3.4

Overall dimensions (WxDxH) cm

50x59x64

50x59x64

Weight

kg

103

130



All measurements have been carried out at: rated voltage and frequency

ambient temperature: 20 °C

Technical changes without prior notification reserved.