JULABO LH 85 High Dynamic Temperature System User Manual

Page 68

Advertising
background image

Remote control

68

11.3. List of commands


When the RS485 interface is used, the instrument address stands in front of each command (Axxx_).

in commands: Asking for parameters or temperature values to be displayed.

Command

Parameter

Response of instrument

version

none

Number of software version (V X.xx)

status

none

Status message, error message (see page 71)

in_pv_00

none

Actual bath temperature.

in_pv_01

none

Heating power being used (%).

in_pv_02

none

Temperature value registered by the external Pt100 sensor.

in_pv_03

none

Temperature value registered by the safety sensor. >TANK<

in_pv_05

none

Pump pressure in bar.

in_sp_00

none

Working temperature

in_sp_03

none

High temperature warning limit

in_sp_04

none

Low temperature warning limit

.

in_sp_05

none

Setpoint temperature of the external programmer
(socket 3.1 - REG+E-PROG) .

in_sp_07

none

Pump pressure stage

in_hil_00

none

Max. cooling power (%).

in_hil_01

none

Max. heating power (%).

in_mode_02

none Identification

type:

0 = no identification

1 = single identification

2 = continual identification

in_mode_03

none

Type of the programmer input:

0 = Voltage 0 V to 10 V

1 = Current 0 mA to 20 mA

in_mode_04

none

Internal/external temperature control:

0 = Temperature control with internal sensor.

1 = Temperature control with external Pt100 sensor.

Advertising
This manual is related to the following products: