8.8. Sensors



ATC - Absolute Temperature Calibration

 Select the submenu "Temp.Sensor" with enter

.

 Select the desired option with the up/down cursor keys

.

A flashing digit indicates that a value needs to be

entered i.e. set.
ATC Int: internal

sensor

ATC Ext: external

sensor

 Enter the desired compensating value and set this value by

pressing enter

.


Notice:

ATC Int: Do not alter the factory-setting!


ext. Pt100


Calibrated thermometer (T

M

)

Presto: Pt100 sensor (T

ExtIst

)

Setp.: xx.00°C
IntAct xx.00°C
ExtAct: 51.22°C
Control: Intern

ATC Ext:

serves calibration of an external Pt100 sensor.

 Immerse the external Pt100 sensor and a calibrated

thermometer together in a calibration bath at 50°C, for example,
and allow the reading to stabilize.

 Read the temperatures of the calibrated thermometer (T

M

) and

the external Pt100 sensor (T

ExtIst

) in their respective displays.

 Calculate the temperature difference and enter and set this

difference value as the ATC external compensating parameter.
(Example:

T = T

ExtIst

- T

M

= 1.22 K)

 Use the keypad to enter the desired compensating value

(e.g. 1.22 °C) and set with enter

.

Important:

The ATC function remains active until reset to 00.00 °C.