8.5. Analog

inputs/outputs









or



This submenu enables setting of the input and output values for
the programmer input and the temperature recorder outputs of
socket REG+E-PROG (21).

 Press enter

to select the inputs/outputs submenu.

 Use the up/down cursor keys

to select the desired

option and press enter

to open.

Chan.1

voltage output for recorder (V)

Chan.2

voltage output for recorder (V)

Chan.3

current output for recorder

(mA)

EPROG

external programmer input

 First define the desired output value for channels 1 to 3:

Press the P-key

to select the desired output value and

set with enter

.


Setpoint active setpoint temperature

(Keypad, integr. programmer/ext. programmer)

ActInt

internal actual temperature value

(bath

temperature)

ActExt

external actual temperature value

(external

sensor)

Power

periodic or intermittent heating or cooling

 Then select the display size for channels 1 to 3:


Channel 1 and 2 voltage outputs
Assign the voltage values of 0 V to the lowest and 10 V to the
highest necessary temperature or power rating required as an
output value (°C / %).
Current output channel 3
Assign the current values 0 mA or 4 mA to the lowest and 20
mA to the highest temperature or power rating required as an
output value (°C / %).

The current output offers 2 ranges for selection:
0 to 20 mA and 4 to 20 mA

Select the desired range by pressing the P-key

and set with enter

.

The LCD display changes automatically.