JULABO LH 85 High Dynamic Temperature System User Manual

Page 42

Advertising
background image

Menu functions

42

8.1. Configuration


By means of the configuration functions, operation of the
instrument can be optimized for the current application.

 Press enter

to select the configuration submenu.

 Use the up/down cursor keys

to select the desired

option. A flashing line indicates that a value needs to be
entered.

 Press the P-key

to select the parameter and press enter

.

 Press escape

to return the previous menu level.



> Identification

When performing an identification for the controlled system
(temperature application system), the control parameters Xp, Tn
and Tv will be automatically determined and stored.

Possible parameters:

off - no identification.

The control parameters ascertained during the last
identification are used for control purposes.

once - single identification

The instrument performs a single identification of the
controlled system after start.
After the identification process the parameter is automatically
set to "off".

always - continual identification

The instrument performs an identification of the controlled
system whenever a new setpoint is to be reached.
NOTE: Use this setting only when the temperature application
system changes permanently.

Important:

For physical reasons, refrigeration units provide full cooling
performance only after some minutes. During identification this is
given consideration by waiting for 5 minutes after switch-on.

Advertising
This manual is related to the following products: