JULABO LH 85 High Dynamic Temperature System User Manual

Page 70

Advertising
background image

Remote control

70

Command

Parameter

Response of instrument

out_mode_04

0

Temperature control of internal bath.

out_mode_04

1

External control with Pt100 sensor.

out_mode_05

0

Stop the unit = R –OFF-.

out_mode_05

1

Start the unit.

out_mode_08

0

Set the control dynamics - aperiodic

out_mode_08

1

Set the control dynamics - standard

out_sp_00

xxx.x

Set working temperature.

out_sp_03

xxx.x

Set high temperature warning limit

.

out_sp_04

xxx.x

Set low temperature warning limit

.

out_sp_07

x

Set the pump pressure stage (1 ... 5).

out_hil_00

-xxx

Set the desired maximum cooling power (0 % to 100 %).
Note: Enter the value with a preceding negative sign!

out_hil_01

xxx

Set the desired maximum heating power (10 % to 100 %).

out_par_04

xxx

Band limiting during external control.
Setting the maximum difference between the temperatures
in the internal bath and external system.

out_par_06

xxx

Xp control parameter of the internal controller.

out_par_07

xxx

Tn control parameter of the internal controller.

out_par_08

xxx

Tv control parameter of the internal controller.

out_par_09

xxx

Xp control parameter of the cascade controller.

out_par_10

xxx

Proportional portion of the cascade controller.

out_par_11

xxx

Tn control parameter of the cascade controller.

out_par_12

xxx

Tv control parameter of the cascade controller.

out _par_13

xxx

Maximum internal temperature of the cascade controller.

out _par_14

xxx

Minimum internal temperature of the cascade controller.