JULABO LH 85 High Dynamic Temperature System User Manual

Page 72

Advertising
background image

Remote control

72

-14 TEMPERATURE/LEVEL ALARM

Safety temperature alarm
Low liquid level alarm

-15 EXTERNAL SENSOR ALARM

External control selected, but external Pt100 sensor
not connected.

-16 TRIAC/RELAY CONNECTION OPEN

Heating circuit interrupted.

-17 TRIAC SHORTED

Heating circuit short-circuited.

-20 WARNING: CLEAN CONDENSOR OR

CHECK COOLING WATER CIRCUIT
OF REFRIGERATOR

Cooling of the condenser is affected.
Clean air-cooled condenser.
Check the flow rate and cooling water temperature on
water-cooled condenser.

-21 WARNING: COMPRESSOR STAGE 1

DOES NOT WORK

Compressor stage 1 does not work.
Check mains circuit breakers F1 and F3, both might
have triggered.
Reset circuit breakers if necessary.

-26 WARNING: STAND-BY PLUG IS

MISSING

External stand-by contact is open.
Stand-by input - see pages 44 and 6.

-31 INTERNAL COMMUNICATION

ERROR

Communication between remote module RD and
instrument interrupted.

-40 NIVEAU LEVEL WARNING

Low liquid level warning in the internal reservoir.

-41 HIGH LIQUID LEVEL WARNING

High liquid level warning in the internal reservoir.