JULABO LH 85 High Dynamic Temperature System User Manual

Page 36

Advertising
background image

Manual operation

36

7.2. Direct setting of the working temperature

Setp.: 20.00°C
IntAct 21.00°C
ExtAct: --.--°C
Control: Intern

This setting may be carried out with the temperature system being
in operating state Start or Stop!


 The value previously set appears on the DIALOG-DISPLAY

(LCD) (example: 20.00 °C).
A flashing segment indicates that a value needs to be entered.

 Use the keypad to enter the new value and press enter

to store the selected value.

(examples: 120.00 °C; 228.50 °C; -15.00 °C).


7.3. Settings in the SET menu

 Press the key

to call up the SET menu. It is then

indicated on the DIALOG-DISPLAY (LCD).

Set the following values:
in Line 1: the working temperature in °C

Setp.:

Line 2: the high temperature limit in °C

Overt.:

Line 3: the low temperature limit in °C

Subtmp:

Line 4: the pump pressure stage

St.Pump:

 Enter the desired values via the keypad. Settings may be

carried out with the temperature system being in operating
state Start or Stop!

 Use the cursors

to swap the lines.

 Press

escape

to quit the SET menu.


7.3.1.

Setting the working temperature

>Setp.: 200.00°C
Overt.: 255.00°C
Suptmp: -55.00°C
St.Pump: 2

>Setp.: 120.00°C
Overt.:255.00°C
Subtmp:-55.00°C
St.Pump: 2

>Setp:

Setting the working temperature

Setting in line 1.

The value previously set appears on the DIALOG-DISPLAY
(LCD) (example: 100.00 °C).
A flashing segment indicates that a value needs to be
entered.

Use the keypad to enter the new value

(examples: 120.00 °C; 228.50 °C; -15.00 °C).

Press enter

to store the selected value.

Advertising