JULABO LH 85 High Dynamic Temperature System User Manual

Page 76

Advertising
background image

Cleaning the unit

76


Cleaning is performed in two steps.
First rinse the internal reservoir, the tubing lines and the heat
exchanger.
As second step, carefully remove the cleaning liquid.

Pump
>Pump OFF
Mode: Sys close

Pump
>Pump OFF
Mode: drain

1. Cleaning internal reservoir, tubing lines and heat exchanger:

 Fill the temperature system with cleaning liquid.

See chapter 5.7. Filling

 In the submenu >Pump< the unit has to be adjusted on the

operating status >Sys close< .

 Set the working temperature setpoint.

Recommendation: 20 °C

 Press the start key

and allow the liquid to be pumped through

the unit for about 15 minutes.

 Press the stop key

.

Drain the temperature system.
In the submenu >Pump< adjust the menu option >Mode< on
>drain< stellen.

See chapter 5.9. Draining

2. Removing the cleaning liquid:
Switch on the unit with the mains switch!

The venting valve remain open.

 Blow through the temperature system with dry, warm air.

The air is fed in turns through the different connectors on the
temperature system.
Begin with the connectors (15, 16) leading to the internal
reservoir.
Then blow through the pump connectors (13, 14) leading to the
heat exchanger and the drain port (8) on the front.

 Allow the system to dry (wait until the humidity evaporates). Then

reconnect the temperature system or close the connectors
according the intended use.
Tighten the drain screw on the front.

Advertising
This manual is related to the following products: